Indian Journal of Critical Care Medicine
Medknow Publications on behalf of the Indian Society of Critical Care Medicine
ISSN: 0972-5229 EISSN: 1998-359x
Vol. 12, Num. 1, 2008, pp. 10-14

Indian Journal of Critical Care Medicine, Vol. 12, No. 1, January-March, 2008, pp. 10-14

Review Article

Optimal positioning of right-sided internal jugular venous catheters: Comparison of intra-atrial electrocardiography versus Peres' formula

Department of Anesthesia and Critical Care, GR Doshi and KM Mehta Institute of Kidney Diseases and Research Centre, Dr. HL Trivedi Institute of Transplantation Sciences, Civil Hospital campus, Asarwa, Ahmedabad - 380 016, Gujarat
Correspondence Address:49/B 'NAYANRAMYA', Chaitanyanagar, Nr. Stadium Petrol Pump, Stadium Road, P.O. Navjivan, Ahmedabad - 380 014, Gujarat
dranishjoshi@yahoo.com

Code Number: cm08002

Abstract

Central venous catheters are routinely placed in patients undergoing major surgeries where expected volume and hemodynamic disturbances are likely consequences. The incorrect positioning may give false central venous pressure (CVP) readings leading to incorrect volume replacement and other serious complications. 50 American Society of Anaesthesiologists grade II-IV patients aged 18-60 years were selected for right-sided internal jugular vein (IJV) catheterization using Seldinger's technique. In group A, central venous catheterization was done under electrocardiography (ECG) guidance. In group B, the catheter was inserted blindly using Peres' formula of "height (in cm)/10". The position of the tip of central venous catheter was confirmed radiologically by postoperative chest X-ray. 92% of patients in group A had radiologically correct positioning of catheter tip i.e. above the carina, while in group B 48% patients had over-insertion of the catheter in to the right atrium. Intra-atrial ECG technique to judge correct tip positioning is simple and economical. It can determine the exact position intraoperatively and can justify a delayed postoperative chest X-ray to confirm CVC line tip placement.

Keywords: Central venous catheterization, intra-atrial electrocardiography, Peres′ height formula

Introduction

Central venous catheters (CVC) are routinely placed in patients undergoing major surgeries where expected volume and hemodynamic disturbances are likely consequences, and/or for multiple infusion administration during surgery or later for parenteral nutrition postoperatively. The correct position of the tip of CVC is considered to be in the superior vena cava (SVC) above the level of pericardial reflection. Blood flow conditions are then optimal to keep the catheter away from the intima and to dilute the infused drugs immediately. Anaesthesiologists usually place the lines in the operating room preoperatively and the chest X-ray is done postoperatively to confirm the correct placement several hours later. Because incorrect placement can lead to serious complications such as cardiac tamponade, perforation or dysrhythmias caused by interaction with the vessel wall or the endocardium, [1] it would be desirable to confirm correct positioning after placement in the operating room. The best way to confirm correct position is to perform the procedure under fluoroscopy or to obtain a chest X-ray postprocedure, but it is costly, associated with radiation exposure and would lengthen the operative time.

Various landmarks, [2],[3] simple formulae [4],[5] and sophisticated techniques like right atrial ECG [6],[7] and transesophageal echocardiography [7],[8] have been developed to ensure correct placement of the CVC tip. Peres′ utilized the patient′s height to predict the optimal length of the catheter to be inserted by different approaches and demonstrated that 24% terminated in right atrium. [9] Another technique is to place the CVC under ECG guidance which was first described by Hellerstein and colleagues in 1949 [10] and has been claimed to detect the intraatrial position of the tip of CVCs by detecting an intraatrial P-wave (P-atriale) with the exploring electrode.

The aim of this study was to compare the intra-atrial ECG technique with Peres′ height formula for optimal placement of CVC and establish it′s accuracy by postoperative chest X-ray.

Materials and Methods

The protocol was approved by the institutional ethics committee and informed consent were obtained from 50 patients aged 18-60 years undergoing elective surgery and scheduled to receive a CVC as part of their anaesthetic management were enrolled in this prospective study. Height of all patients in cms was recorded at the time of a preoperative visit, and the patients were randomly divided into 2 groups of 25 each. Patients with altered coagulation parameters or having cardiac disorders like atrial fibrillation, multifocal ventricular premature beats, supraventricular tachycardia, left bundle branch block or pacemaker were excluded from the study.

In both the groups, after successful puncture of the right internal jugular vein (IJV), a 15/20 cm single lumen CVC (Certofix ® Mono 16G, B. Braun Melsungen, Germany) was inserted over 50 cm guidewire using the Seldinger technique. In Group A, the guidewire was withdrawn through the catheter until a mark on the guidewire indicated the tip to be exactly positioned at the tip of the catheter. A connection between the guidewire and the ECG adapter (Certodyn ® - Universal adapter, B. Braun Melsungen, Germany, Cost Rs. 9000.00) was established by connecting the reference red electrode to a universal adapter equipped with a switching function on the right thoracic side to record a modified lead II. The yellow electrode was placed on the left shoulder and the neutral green electrode was placed on the lower left chest. By turning the switch of the universal adapter, intra-atrial ECG could be recorded [Figure 1A] and [Figure 1B].

While advancing the guidewire alongwith the catheter, the configuration of the P wave was seen on the ECG monitor. Gradually the height of the P wave increased and it became equal to or more in amplitude than the R wave [Figure 2A]. On further advancement, the P wave became bifid, but at this point the catheter and guidewire were immediately withdrawn till the P wave became of the normal configuration [Figure 2B]. The CVC was fixed with stitches and sterile dressing was applied.

In Group B, the catheter was blindly inserted and final insertion depth was kept as per the Peres′ formula of "height (in cm)/10". [4] A postoperative portable chest X-ray anteroposterior was done in all cases in supine position for assessing the position of the catheter′s tip. Chest X-rays were read by one radiologist who was aware of the study protocol but blinded to the group allocation. CVC position was judged as correct if the tip was positioned above the level of carina and as overinsertion when the tip was below the level of carina. In both the groups, final insertion depth, incidence of arrhythmias during the CVC placement, insertion time and complications (arterial puncture, hematoma, pneumothorax) were recorded.

Data and Statistical Analysis

Demographic data are presented as mean ± SD. Correct CVC position, incidence of arrhythmias and complications are presented as percentages. Data were analyzed using the unpaired Student′s t test. P < 0.001 was considered to be highly significant.

Results

Both groups were comparable with respect to age, body weight, sex and height [Table - 1].

The mean insertion time was 6.5 min in Group A as compared to 5 min in Group B. No arrhythmias were noted in Group A, while 3 of 25 patients (12%) had dysrhythmias in Group B on insertion which was corrected immediately by pulling back the catheter by 2 cm. The mean length of the IJV catheter inserted in Group A was 12.24 ± 1.30 cm compared to 15.88 ± 0.88 cm in Group B, the difference of 3.64 cm was highly significant ( P < 0.001) [Table - 2].

On postoperative chest X-ray in 23 of 25 (92%) cases in Group A the IJV catheter was properly positioned, while in 12 of 25 cases (48%) there was overinsertion in Group B. Also there was coiled catheter in 6 of 25 cases (24%) in Group B while in 2 of 25 cases (8%) in Group A [Table - 3].

Discussion

Depending on the point of entry or indication for a CVC, the optimal site for CVC tip placement varies. [11] However, guidelines for CVC placement recommend that the catheter tip should lie above the pericardial reflection [3],[9],[11],[12],[13] to prevent serious and potentially lethal complications like cardiac tamponade, malignant arrhythmias, placement in coronary sinus and tricuspid valve damage. The upper limit of the pericardial reflection cannot be seen on a plain chest X-ray, but it is generally accepted to be below the carina. This has been assessed in preserved [2] and fresh cadavers, [14] in anesthetized children undergoing cardiac surgery [15] and in adults using computerized tomograms. [16] Moreover, its location is preserved even in pulmonary pathology due to its fixation with connective tissue, parallax effect is limited due to its central location and small sagittal distance between it and SVC and it is easily visible even in a poor quality portable anteroposterior chest X-ray. Hence we considered carina as a radiological landmark for CVC tip position.

Various methods have been suggested to estimate the expected length of CVC at the time of insertion. Pere P.W. studied correlation between the length of catheter inserted and patient′s height and observed that catheters inserted through right IJV from midcervical point or lower puncture to Height/10cm ended in SVC, while those inserted more than Height/10 + 1cm, 47% ended in right atrium. [9] In our study, we observed 48% incidence of overinsertion with Height/10cm formula with insertion at cricoid level. This may be because we considered carina as reference point for correct positioning whereas Pere PW. [9] considered SVC/RA junction as optimal tip positioning.

Good quality intra-atrial ECG and clear display of the P-wave on the ECG monitor are essential for successful guidance and positioning of CVCs. Consequently, a limitation of ECG guidance is that it cannot be reliably used in patients with atrial fibrillation or other supraventricular arrhythmias. Analysis of P wave morphology as a marker of ECG-guided central venous catheterization has been evaluated previously and most of these studies report variable results. Jeon et al. , [7] undertook a prospective descriptive clinical study using transoesophageal echocardiography (TEE) guidance to establish the absolute locations and ranges of CVC tip positions when specific P wave patterns are displayed during ECG-guided central venous catheterization. They concluded that the tallest peaked P wave may be used to place the CVC tip at the SVC/RA junction, the normally-shaped P wave identifies the mid to upper SVC and a biphasic pattern of the P wave can be used to locate the RA. Based on this, we considered normally-shaped P waves as correct positioning and avoid overinsertion in ECG guided group.

Previous studies have indicated the usefulness of ECG guided positioning of CVC placement. [6],[17] In a prospective, controlled, randomized trial comparing ECG guidance to conventional placement, Francis et al. , [18] reported a significantly higher success rate (96% v/s 59%) using a continuous column of normal saline for ECG-guided CVC placement. Gebhard RE et al. , reported 96% correct positioning in group ECG and 76% in group NO-ECG. [19] In our study, success rate was 92%, however we could not prevent malposition in two patients but identified it as decrease rather than increase in the height on further advancement of the CVC which was confirmed on postoperative chest X-ray. We did not try repositioning maneuvers in these cases but it could have resulted in correction of this misplacement and higher incidence of success. CVC placement in the RA occurred only in group B and was associated with a mean insertion depth of 15.88 ± 0.88 cm. Based on this, we recommend inserting CVCs via the Right internal jugular vein (RIJV) no deeper than Height/10-2 cm if ECG guidance cannot be used.

One of the limitations of this study is that all CVCs were placed through the RIJV which is used more frequently than the left internal jugular vein (LIJV) as the access vessel for CVC placement. Therefore, no conclusion can be drawn whether ECG guidance results in more accurate positioning when either the LIJV or one of the subclavian veins are chosen as access routes to the SVC. Schummer et al. , [20] suggested that ECG guidance does not improve the accuracy of the CVC placed through the LIJV as it is not capable of providing information about angle of the CVC tip in relation to the wall of the SVC. However, it is believed that ECG guidance is efficient in positioning CVCs into a major vessel just proximal to the heart.

Conclusion

Intra-atrial ECG guided placement is a useful tool for accurate positioning of RIJV-CVC. The procedure is reliable compared to height formula and requires universal adaptor which is not expensive. It can avoid radiologic confirmation in uncomplicated insertions minimizing cost and radiologic exposure to patients and health care workers.

References

1.Defalque RJ, Campbell C. Cardiac tamponade from central venous catheters. Anesthesiology 1979;50:249-52.  Back to cited text no. 1  [PUBMED]  
2.Schuster M, Nave H, Piepenbrock S. The carina as a radiological landmark in central venous catheter placement. Br J Anaesth 2000;85:192-4.  Back to cited text no. 2    
3.Stonelake PA, Bodenham AR. The carina as a radiological landmark for central venous catheter tip position. Br J Anaesth 2006;96:335-40.  Back to cited text no. 3  [PUBMED]  [FULLTEXT]
4.Czepizak CA, O'Callaghan JM, Venus B. Evaluation of formulas for optimal positioning of central venous catheters. Chest 1995;107:1662-4.  Back to cited text no. 4  [PUBMED]  [FULLTEXT]
5.Chalkiadis GA, Goucke CR. Depth of central venous catheter insertion in adults: an audit and assessment of a technique to improve tip position. Anesth Intensive Care 1998;26:61-6.  Back to cited text no. 5    
6.Wilson RG, Gaer JA. Right atrial electrocardiography in placement of central venous catheters. Lancet 1988;27:462-3.  Back to cited text no. 6    
7.Jeon Y, Ryu HG, Yoon SZ, Kim JH, Bahk JH. Transesophageal echocardiographic evaluation of ECG-guided central venous catheter placement. Can J Anesth 2006;53:978-83.  Back to cited text no. 7  [PUBMED]  [FULLTEXT]
8.Andropoulos DB, Stayer SA, Bent ST, Campos CJ, Alvarez M, Fraser CD. A controlled study of transesophageal echocardiography to guide central venous catheter placement in congenital heart surgery patients. Anesth Analg 1999;89:65-70.  Back to cited text no. 8  [PUBMED]  [FULLTEXT]
9.Peres PW. Positioning central venous catheters: A prospective survey. Anesth Intensive Care 1990;18:536-9.  Back to cited text no. 9    
10.Hellerstein HK, Pritchard WH, Lewis RL. Recording of intracavity potentials through a single lumen, saline filled cardiac catheter. Proc Soc Exp Med 1949;71:58-60.  Back to cited text no. 10    
11.Fletcher SJ, Bodenham AR. Safe placement of central venous catheters: Where should the tip of the catheter lie? Br J Anaesth 2000;85:188-91.  Back to cited text no. 11  [PUBMED]  [FULLTEXT]
12.Rutherford JS, Merry AF, Occleshaw CJ. Depth of central venous catheterization: An audit of practice in a cardiac surgical unit. Anaesth Intensive Care 1994;22:267-71.  Back to cited text no. 12  [PUBMED]  
13.Sundaram R, Koteeswaran SK, Smart NG, Stonelake P, Bodenham A. Safe placement of central venous catheters. Br J Anaesth 2006;97:584-5.  Back to cited text no. 13  [PUBMED]  [FULLTEXT]
14.Albrecht K, Nave H, Breitmeier D, Panning B, Tr φger HD. Applied anatomy of the superior vena cava-the carina as a radiological landmark to guide central venous catheter placement. Br J Anaesth 2004;92:75-7.  Back to cited text no. 14    
15.Yoon SZ, Shin JH, Hahn S. Usefullness of the carina as a radiological landmark for central vanous catheter placement in paediatric patients. Br J Anaesth 2005;95:514-7.  Back to cited text no. 15    
16.Caruso LJ, Gravenstein N, Layon AJ. A better landmark for positioning a central vanous catheter. J Clin Monit Comput 2002;17:331-4.  Back to cited text no. 16    
17.Watters VA, Grant JP. Use of electrocardiogram to position right atrial catheters during surgery. Ann Surg 1997;225:165-71.  Back to cited text no. 17  [PUBMED]  [FULLTEXT]
18.Francis KR, Picard DL, Fajardo MA, Pizzi WF. Avoiding complications and decreasing costs of central venous catheter placement electrocardiographic guidance. Surg Gynecol Obstet 1992;175:208-11.  Back to cited text no. 18  [PUBMED]  
19.Gebhard RE, Szmuk P, Pivalizza EG. The Accuracy of electrocardiogram controlled central line placement Anesth Analg 2007;104:65-70.  Back to cited text no. 19    
20.Schummer W, Herrmann S, Schummer C. Intra-atrial ECG is not a reliable method for positioning left internal jugular vein catheters. Br J Anaesth 2003;91:481-6.  Back to cited text no. 20    

Copyright 2008 - Indian Journal of Critical Care Medicine


The following images related to this document are available:

Photo images

[cm08002f1.jpg] [cm08002t1.jpg] [cm08002t3.jpg] [cm08002f2.jpg] [cm08002t2.jpg]
Home Faq Resources Email Bioline
© Bioline International, 1989 - 2024, Site last up-dated on 01-Sep-2022.
Site created and maintained by the Reference Center on Environmental Information, CRIA, Brazil
System hosted by the Google Cloud Platform, GCP, Brazil