search
for
 About Bioline  All Journals  Testimonials  Membership  News


Indian Journal of Critical Care Medicine
Medknow Publications on behalf of the Indian Society of Critical Care Medicine
ISSN: 0972-5229 EISSN: 1998-359x
Vol. 13, Num. 2, 2009, pp. 92-93

Indian Journal of Critical Care Medicine, Vol. 13, No. 2, April-June, 2009, pp. 92-93

Short Communication

U turn to venous air embolism

Department of Anaesthesiology, Dayanand Medical College and Hospital, Ludhiana, Punjab
Correspondence Address:Dept of Anaesthesiology, Dayanand Medical College and Hospital, 48 lal bagh Ludhiana 142 021
harsimrandmc@gmail.com

Code Number: cm09020

PMID: 19881190
DOI: 10.4103/0972-5229.56055

Abstract

There is a definitive risk of venous air embolism when the fluid infusion is complete and the drip set is still open in a glass bottle.We have devised a novel way of preventing the chances of air embolism when the fluid in the glass bottle finishes. It really gives a "U" turn to the chances of venous air embolism.

Keywords: Glass bottles, IV fluid therapy, venous air embolism

Introduction

Air embolism can complicate peripheral IV fluid therapy [1] or central venous catheter monitoring, including problems with IV infusion pumps, [2] improper flushing of IV sets, [3] incorrect injection of drugs into the infusion system, [3] and accidental disconnection of the hub or removal of central venous catheters. [4] The open to air system in glass bottles is also a potential hazard for life-threatening complications.

Air embolism can cause blockage of small vessels in the pulmonary vasculature with compromise of gas exchange, cessation of ventricular pumping caused by blockage of air and arrhythmia, [5] and paradoxical air embolism to the systemic circulation through a probe-patent foramen ovale. [6]

Glass bottles are widely used for administration of intravenous fluids. Various drugs e.g., mannitol are also available only in glass bottles in our part of the world. For the administration of fluids from glass bottles, a needle has to be inserted, which acts as an airway [Figure - 1] and the fluid comes out through the infusion tubing on the basis of the beer can principle. The rate is controlled by a plastic regulating clamp.

Many times the fluid in the bottle finishes, but the regulating clamp is still open. The needle airway is in place and the infusion line eventually becomes open to air, which can lead to fatal air embolism.

For air embolism to occur through an intravenous drip set, there are two prerequisites. First, is that the patient is connected to an intravenous line that is open to air and second is that the intravenous tubing should be placed into a vein whose pressure is below the atmospheric pressure.

Sub atmospheric pressure can develop in a vein when the intravenous tubing is attached to the central line, or the vein is on the upper arm in the lateral position. [7] A right-sided central line in the left lateral position and the patient attempting to breathe against a closed glottis are two other potential conditions of air embolism; there are high chances of sub atmospheric pressure developing in the vein.

There is a definitive risk of air embolism when the fluid infusion is complete and the drip set is still open.

We have devised a novel way of preventing the chances of air embolism when the fluid in the glass bottle finishes. Weix the infusion tubing by giving it a "U" turn so that the loop is below the level of the vein into which the infusion set is attached [Figure - 2]. This ensures that some fluid remains in the loop and an air embolism is prevented. The presence of a U turn loop increases the negative pressure that would be required to aspirate air. A simple maneuver acts as a safety mechanism from a potentially life-threatening complication and decreases the chances of an air embolism.

References

1.Groell R, Schaffler GJ, Rienmueller R. The peripheral intravenous cannula: a cause of venous air embolism. Am J Med Sci 1997;314:299 -302.  Back to cited text no. 1    
2.Rothenberg F, Schumacher JR, Rosenthal RL. Near-fatal pulmonary air embolus from presumed inadvertent pressure placed on a partially empty plastic intravenous infusion bag. Am J Cardiol 1994;73:1035- 6.  Back to cited text no. 2    
3.Levy I, Mosseri R, Garty B. Peripheral intravenous infusion: Another cause of air embolism. Acta Paediatr 1996;85:385- 6.  Back to cited text no. 3    
4.Ross SM, Freedman PS, Farman JV. Air embolism after accidental removal of intravenous catheter. Br Med J 1979;1:987.  Back to cited text no. 4    
5.Adornato DC, Gildenberg PL, Ferrario CM, Smart J, Frost EA. Pathophysiology of intravenous air embolism in dogs. Anesthesiology 1978;49:120 -7.  Back to cited text no. 5    
6.Colohan AR, Perkins NA, Bedford RF, Jane JA. Intravenous fluid loading as prophylaxis for paradoxical air embolism. J Neurosurg 1985;62:839-42.  Back to cited text no. 6    
7.Plumer's Principles and Practice of Intravenous Therapy. Sharon Weinstein, Ada Lawrence Plumer. Lippincott Williams and Wilkins, 2006;163-73.  Back to cited text no. 7    

Copyright 2009 - Indian Journal of Critical Care Medicine


The following images related to this document are available:

Photo images

[cm09020f1.jpg] [cm09020f2.jpg]
Home Faq Resources Email Bioline
© Bioline International, 1989 - 2024, Site last up-dated on 01-Sep-2022.
Site created and maintained by the Reference Center on Environmental Information, CRIA, Brazil
System hosted by the Google Cloud Platform, GCP, Brazil