Journal of Cancer Research and Therapeutics, Vol. 4, No. 1, January-March, 2008, pp. 18-20 Original Article Execution of mantle field with multileaf collimator: A simple approach Prabhakar Ramachandran, Kunhi Parambath P Haresh, Pappiah S Sridhar , Macharla A Laviraj, Pramod K Julka, Goura K Rath Department of Radiation Oncology, All India Instiute of Medical Sciences, New Delhi - 110 029 Code Number: cr08005 Abstract Background: Until very recently mantle field radiotherapy remained the gold standard for the treatment of favorable early-stage Hodgkin's lymphoma. The classic mantle includes all the major lymph nodes above the diaphragm and extends from the inferior portion of the mandible to the level of the insertion of the diaphragm.
Keywords: Hodgkin's lymphoma, mantle field, multileaf collimator, radiotherapy Introduction Hodgkin's lymphoma (HL) was initially described by Thomas Hodgkin's in 1832. [1] In those days, it used to be a dreaded disease. With the evolution of newer techniques in radiotherapy after 1960 and the availability of newer chemotherapeutic drugs, HL has become a prime example of a curable malignancy. Most of the credit for the initial developments in radiotherapy in the treatment of this disease go to Henry Kaplan of Stanford University. [2] Until very recently, mantle field radiotherapy remained the gold standard for the treatment of favorable, early-stage HL. Mantle field planning used to be a cumbersome process, involving repeated simulations to make customized lead shielding blocks for the lungs, humeral head, larynx, etc. It often took a week to finalize the plan. Nowadays, with the wide availability of the multileaf collimator (MLC) inbuilt in the linear accelerator treatment head, this process is much easier. This paper describes a simple approach in defining the mantle field with MLC by splitting the anterior and posterior field into multiple subfields. Materials and Methods The patient was immobilized with Vacloc and uniframe head mask (Med-Tec) in the supine position and then scanned in a Somatom Volume Zoom CT scanner for 2.5 mm slice thicknesses. The CT datasets were transferred to the Eclipse treatment planning system through the DICOM network for treatment planning. Bones were delineated using the segmentation wizard available in the treatment planning system. The nodal regions were contoured on individual slices for fitting the MLCs. With the help of bony landmarks and the planning target volume (PTV), the MLC shapes were designed to obtain the shape of the mantle field. Elekta Precise equipped with 40 pairs of MLC was used and the whole treatment was performed with 6 MV x-rays. The anterior mantle field was divided into three separate MLC segments. The collimator was kept at 0°. The first MLC segment was shaped to cover the neck, clavicular regions, and mediastinum. The MLC leaves were moved in or out to get the required shape. The second and the third MLC segments covered the right and left axilla, respectively. The posterior field was designed in a similar fashion. Three reference points were defined in the mid-plane for the three MLC subfields and the dose was prescribed to their respective reference points. The whole treatment was performed in the SAD technique. For inhomogeneity correction, equivalent tissue-air ratio method was used for dose calculation. Summation of the three anterior subfields and their matching was verified by exposing an enhanced dose range (EDR) film.Results [Figure - 1][A-C] shows the anterior MLC subfields. [Figure - 1]A shows the first MLC segment shaped to cover the neck, clavicular regions, and mediastinum. [Figure - 1]B and C show the MLC segment shaped to cover the right and left axilla, respectively. Similarly, [Figure - 2] [A-C] shows the posterior MLC subfields opposed to their respective anterior subfields. [Figure - 3] shows the resultant coronal dose distribution obtained out of the combined anterior and posterior MLC subfields.Discussion Until recently the mantle field technique remained the standard for good prognosis, early-stage HL treated by radiotherapy alone. The present trend in the treatment is to use fewer chemotherapy cycles, along with less dose and volume of radiotherapy, to get a better overall therapeutic ratio. [3] In a recent study, Hoskins et al . [4] have shown that involved-field radiotherapy for stage I and IIA HL resulted in a 11% greater risk of relapse compared with extended-field radiotherapy. Even though mantle is less frequently used nowadays, the planning and execution of mantle still remains a concern for the radiation oncologist, physicist, dosimetrist, and the radiation technologist. There are a number of advantages in adopting the MLC technique instead of using the conventional lead blocks for shaping the mantle field. The conventional technique requires considerable amount of time for preparation of the shielding blocks; this can be avoided with MLCs and the treatment can be started without any delay. The MLC technique also considerably reduces the time required for delivering a fraction as compared to the conventional mode. The MLC-based mantle technique avoids the risk of injury to the patient due to inadvertent falling of shielding blocks from the shadow tray while placing and removing them. It also avoids any chance of mismatch in the placement of the anterior and posterior shielding blocks during treatment. It eliminates the exposure to toxic lead, both for the radiation staff and the patient. In case of conventional mantle field employing blocks, the presence of hotspots are unavoidable inside the treated volume. Using multiple MLC subfields, hotspots arising at areas above the supraclavicular region and other areas where effective anterior/posterior separation are thin can be minimized. This eliminates the use of compensators [5] and also provides better dose homogeneity within the treated volume than the conventional method. With more and more cobalt being replaced with linac, the use of MLCs for mantle field with this technique simplifies the whole treatment process. MacDonald et al . [6] has described a technique for defining the mantle field with MLC by rotating the collimator either to 90° or 270°. The disadvantage here is that we still might need additional blocks to the mantle field to shield the extra lung volume coming within the field above the hilar area. Our method avoids the requirement for any additional shielding blocks other than the use of the MLC. In this method, during matching of the three MLC subfields [Figure - 1] A - C , due to a tongue-and-groove effect, there may be a slight underdosage at the junction, which can be avoided by shifting the junction on alternate fractions. In this study, the prescription dose was kept at 25 Gy, and so we did not use any laryngeal shielding. In situations where the prescription dose is more than 25 Gy, the whole treatment can be divided into phase I and phase II. Phase I can be carried out without any laryngeal shield up to 20 Gy and in phase II, the larynx may be shielded with MLCs. A set of MLCs of one bank can be made to overtravel to the other end, at the level of the larynx, in phase II. Those shielded portions which are adjacent to the larynx on either side can be treated with two separate MLC subfields. With this method, the need for a separate laryngeal shielding block can be avoided. The maximum possible field size in Elekta Precise is 40 x 40 cm at 100 cm SSD. For fields more than 40 cm, the patient can be treated at extended SSD. One might argue about the interleaf leakage with the MLC that may lead to a high dose being delivered to the underlying lung. Here one should not forget that in case of conventional mantle fields, shielding achieved by 5 HVL thickness may also result in a similar dose. The X-jaws of Elekta Precise machine has the capability of over-traveling the central axis by 12.5 cm. So the secondary back-up jaws (X-jaws) can still minimize the interleaf leakages. The combination of MLC and jaws for shielding further reduces the scattered dose to the patient. The only disadvantage of our technique is that it requires multiple subfields for achieving the conventional mantle dose distribution. Conclusion The technique described in this study is very simple, easy to implement, and avoids unnecessary delay in the execution of the mantle field. The shape of the mantle block can be easily achieved with the MLCs.References
Copyright 2008 - Journal of Cancer Research and Therapeutics The following images related to this document are available:Photo images[cr08005f3.jpg] [cr08005f1.jpg] [cr08005f2.jpg] |
|