Noninvasive monitoring of PaCO2 during one-lung ventilation and minimal access surgery in adults: End-tidal versus transcutaneous techniques

Department of Anesthesiology and Pediatrics, University of Missouri, Columbia, Missouri
Correspondence Address:University of Missouri, Department of Anesthesiology, 3W-27G HSC One Hospital Drive, Columbia, Missouri 65212, tobiasj@health.missouri.edu

Code Number: ma07003

Abstract

Background: Previous studies have suggested that end-tidal CO 2 (ET-CO 2 ) may be inaccurate during one-lung ventilation (OLV). This study was performed to compare the accuracy of the noninvasive monitoring of PCO 2 using transcutaneous CO 2 (TC-CO 2 ) with ET-CO 2 in patients undergoing video-assisted thoracoscopic surgery (VATS) during OLV.
Materials and Methods:
In adult patients undergoing thoracoscopic surgical procedures, PCO 2 was simultaneously measured with TC-CO 2 and ET-CO 2 devices and compared with PaCO 2 .
Results:
The cohort for the study included 15 patients ranging in age from 19 to 71 years and in weight from 76 to 126 kg. During TLV, the difference between the TC-CO 2 and the PaCO 2 was 3.0 ± 1.8 mmHg and the difference between the ET-CO 2 and PaCO 2 was 6.2 ± 4.7 mmHg ( P =0.02). Linear regression analysis of TC-CO2 vs. PaCO 2 resulted in an r 2 = 0.6280 and a slope = 0.7650 ± 0.1428, while linear regression analysis of ET-CO 2 vs. PaCO 2 resulted in an r 2 = 0.05528 and a slope = 0.1986 ± 0.1883. During OLV, the difference between the TC-CO 2 and PaCO 2 was 3.5 ± 1.7 mmHg and the ET-CO 2 to PaCO 2 difference was 9.6 ± 3.6 mmHg ( P =0.03 vs. ET-CO 2 to PaCO 2 difference during TLV; and P <0.0001 vs. TC-CO 2 to PaCO 2 difference during OLV). In 13 of the 15 patients, the TC-CO 2 value was closer to the actual PaCO 2 than the ET-CO 2 value ( P =0.0001). Linear regression analysis of TC-CO 2 vs. PaCO 2 resulted in an r 2 = 0.7827 and a slope = 0.8142 ± 0.0.07965, while linear regression analysis of ET-CO 2 vs. PaCO 2 resulted in an r 2 = 0.2989 and a slope = 0.3026 ± 0.08605.
Conclusions:
During OLV, TC-CO 2 monitoring provides a better estimate of PaCO 2 than ET-CO2 in patients undergoing VATS.

Keywords: End-tidal CO 2 , thoracoscopy, transcutaneous CO 2

The measurement of the partial pressure of carbon dioxide in the arterial blood (PaCO 2 ) is used to determine the adequacy of ventilation and to guide changes in mechanical ventilation. Although arterial sampling provides this information, frequent arterial blood gas (ABG) analysis is expensive, time restrictive and provides only intermittent data for what may be a rapidly changing value. To overcome such problems, noninvasive monitors are frequently used to provide a continuous estimate of the PaCO 2 . In the operating room, end-tidal (ET) CO 2 monitoring remains the standard of care for continuous and noninvasive PaCO 2 monitoring. However, various factors, including patient positioning, sampling errors and ventilation-perfusion mismatch due to dead space or shunt, may significantly affect the arterial to ET-CO 2 gradient.[1],[2],[3]

One-lung ventilation (OLV) is commonly used during thoracic surgery to allow surgical access to the operative lung and to eliminate lung movement during ventilation. The technique is helpful during open thoracic procedures and a necessity for minimal access thoracoscopic procedures. During OLV, the bronchus of the affected lung is occluded by a bronchial blocker or isolated by a double-lumen endotracheal tube, with oxygenation and ventilation supported by the nonoperative lung. Even with effective hypoxic pulmonary vasoconstriction, the technique results in an increase in the shunt fraction and perfusion of the nonventilated lung. Previous studies have demonstrated a significant arterial to ET-CO 2 gradient during one-lung OLV.[4],[5],[6] The efficacy of ET-CO 2 monitoring is limited during such procedures, suggesting that alternative noninvasive monitors of PaCO 2 may be needed. Although used predominantly in the neonatal and pediatric population, there is increasing interest in the use of and recent reports of transcutaneous (TC) CO 2 monitoring in the adult population.[5],[6],[7],[8],[9] The current study prospectively compares ET-CO 2 and TC-CO 2 monitoring during OLV in adults undergoing video-assisted thoracoscopic surgery.

Materials and Methods

The study was approval by the Institutional Review Board and the Committee for the Protection of Human Subjects of the University of Missouri. Verbal informed consent was obtained from each patient. The patient population included patients scheduled for minimal access thoracic surgery and OLV, who were 18 years of age or older and in whom intraarterial access was deemed necessary. OLV was provided by either a bronchial blocker or a double lumen endotracheal tube. Effective lung separation was confirmed by clinical auscultation and fiberoptic bronchoscopy prior to the start of the procedure. Intrathoracic CO 2 insufflation was not used during the procedure.

Transcutaneous and end-tidal carbon dioxide monitoring

ET-CO 2 was measured using an infrared analyzer with a side stream sampler attached at the elbow between the endotracheal tube and the anesthesia circuit. Prior to use, the ET-CO 2 device was calibrated according to the manufacturer's recommendations. TC-CO 2 was measured with a commercially available TC-CO 2 device (Sentec AG, Therwil, Switzerland). This device is based on a Stow-Severinghaus-type CO 2 sensor combined with a pulse oximeter and is attached to the patient's earlobe by a low pressure attachment clip.[10],[11] The in vitro 90% response time is < 50s for the CO 2 electrode. Prior to placement, the sensor was prepared and calibrated according to the manufacturer's recommendations. The sensor is calibrated in vitro by using a one-point dry gas calibration with 7% carbon dioxide. The working temperature of the sensor is 42°C. The sensor was cleaned with alcohol and dried before application. One drop of contact gel was applied to the center of the sensor prior to placement on the patient's earlobe. Oxygen saturation values are available immediately, while TC-CO 2 values are available after a 2-3 min calibration time.

Data collection and statistical analysis

ABG analyses were obtained following endotracheal intubation during two-lung ventilation (TLV) and as clinically indicated during OLV. When an ABG was obtained, the ET-CO 2 and TC-CO 2 were simultaneously recorded on a data sheet. Calculation of the absolute difference between the noninvasive monitor (ET-CO 2 or TC-CO 2 ) and the PaCO 2 was performed. Negative numbers were not used because this could artificially lower the mathematical mean of the differences between the noninvasive monitors of CO 2 and the PaCO 2 . If multiple ABGs were obtained during OLV, the absolute differences between the PaCO 2 and the noninvasive monitors were averaged and counted as a single data point. This was done to avoid biasing the data by overrepresentation of any one patient as the number of ABGs varied for each patient. The absolute difference between the ET-CO 2 and PaCO 2 was compared to the absolute difference between the TC-CO 2 and the PaCO 2 during TLV and OLV using a non-paired t-test. A contingency table with a Fisher's exact test was used to compare the times that each of the noninvasive monitors was closest to the actual PaCO 2 . Using the raw numbers from all of the individual sample sets (PaCO 2 /ET-CO 2 /TC-CO 2 ) obtained during TLV and OLV, linear regression analysis and Bland-Altman analyses were performed.

Results

The cohort for the study included 15 patients ranging in age from 19 to 71 years (46.9 ± 17.3 years) and in weight from 76 to 126 kg (93.4 ± 12.4 kg). There were 11 men and 4 women. Nineteen sample sets (PaCO 2 /ET-CO 2 /TC-CO 2 ) were obtained during TLV and 27 were obtained during OLV. During TLV, the difference between the TC-CO 2 and the PaCO 2 was 3.0 ± 1.8 mmHg with a range of 0 to 8 mmHg and the difference between the ET-CO 2 and the PaCO 2 was 6.2 ± 4.7 mmHg with a range of 2 to 18 mmHg ( P = 0.02 vs. TC-CO 2 to PaCO 2 difference) [Table - 1]. During TLV, the TC-CO 2 value was closer to the actual PaCO 2 in 10 patients, the ET-CO 2 value was closer in 2 patients and there was no difference in 3 patients ( P =NS). During TLV, linear regression analysis of TC-CO 2 vs. PaCO 2 resulted in an r 2 = 0.6280 and a slope = 0.7650 ± 0.1428 (95% confidence intervals: 0.4367 to 1.066). During TLV, linear regression analysis of ET-CO 2 vs. PaCO 2 resulted in an r 2= 0.05528 and a slope = 0.1986 ± 0.1883 (95% confidence intervals: -0.1956 to 0.5928) [Figure - 1]. During TLV, Bland-Altman analysis of TC vs. PaCO 2 revealed a bias of +2.2 mmHg and a precision of ± 3.0 mmHg and analysis of ET vs. PaCO 2 revealed a bias of -7.3 mmHg and a precision of ± 5.5 mmHg.

During OLV, the difference between the TC-CO 2 and the PaCO 2 was 3.5 ± 1.7 mmHg with a range of 2 to 8 mmHg ( P = NS vs. TC-CO 2 to PaCO 2 difference during TLV). During OLV, the ET-CO 2 to PaCO 2 difference was 9.6 ± 3.6 mmHg with a range of 2 to 15 mmHg ( P = 0.03 vs. ET-CO 2 to PaCO 2 difference during TLV; and P < 0.0001 vs. TC-CO 2 to PaCO 2 difference during OLV) [Table - 1]. In 13 of the 15 patients, the TC-CO 2 value was closer to the actual PaCO 2 than the ET-CO 2 value ( P = 0.0001). During OLV, linear regression analysis of TC-CO 2 vs. PaCO 2 resulted in an r 2= 0.7827 and a slope = 0.8142 ± 0.07965 (95% confidence intervals: 0.6513 to 0.9771). During OLV, linear regression analysis of ET-CO 2 vs. PaCO 2 resulted in an r 2 = 0.2989 and a slope = 0.3026 ± 0.08605 (95% confidence intervals: 0.1266 to 0.4786) [Figure - 2]. During OLV, Bland-Altman analysis of TC vs. PaCO 2 revealed a bias of + 2.9 mmHg and a precision of ± 2.5 mmHg and analysis of ET vs. PaCO 2 revealed a bias of -10.5 mmHg and a precision of ± 6.2 mmHg.

No problems occurred with ET-CO 2 monitoring during the study period. In two patients, the TC-CO 2 electrode had to be repositioned after the initial placement to obtain an effective value. No blistering, erythema or skin changes were noted on the earlobe following use of the TC-CO 2 device.

Discussion

The current study demonstrates that noninvasive monitoring of PaCO 2 using ET-CO 2 devices may be inaccurate during minimal access thoracoscopic surgery and OLV in adults. In our cohort of 15 patients, TC-CO 2 monitoring was more accurate than ET-CO 2 monitoring (difference of 3.0 ± 1.8 mmHg vs. 6.2 ± 4.7 mmHg) during the baseline state using TLV. Although no significant change was noted in the TC-CO 2 to PaCO 2 gradient during OLV (3.5 ± 1.7 mmHg), the ET-CO 2 to PaCO 2 difference increased to 9.6 ± 3.6 mmHg. In 13 of 15 patients, the TC-CO 2 was closer to the actual PaCO 2 than the ET-CO 2 during OLV.

Several factors may be responsible for discrepancies between ET-CO 2 and PaCO 2 , including technical issues with the monitor; and patient-related factors, including ventilation-perfusion mismatch, dead space and true shunt.[12],[13] Whitesell et al demonstrated that patients with underlying lung disease had a significantly greater ET-CO 2 to PaCO 2 gradient when compared with patients with normal baseline pulmonary function (3.3 ± 0.6 mmHg versus 0.8 ± 0.3 mmHg).[13] Patient positioning has also been shown to have an impact on the accuracy of ET-CO 2 monitoring.[2],[3] With patients undergoing renal or upper ureteral surgery in the supine position, Pansard et al reported that the ET-CO 2 to PaCO 2 difference was 4.8 ± 3.9 mmHg 10 min after induction and increased to 7.9 ± 3.5 mmHg ( P < 0.01) 5 min after placement of the patients into the lateral decubitus 'kidney rest' position. Similar results were reported by Grenier et al in a cohort of patients undergoing neurosurgical procedures in the lateral decubitus position.[3]

In addition to these factors, the significant physiologic alterations induced by OLV can be expected to alter the relationship between ET-CO 2 and PaCO 2 values. Ip Yam et al evaluated the accuracy of ET-CO 2 during OLV in a cohort of 22 adults undergoing thoracotomy.[4] During TLV, the ET-CO 2 to PaCO 2 difference was 1.3 ± 0.6 kPa (1 kPa = 7.5 mmHg) and it was 1.2 ± 0.7 kPa during OLV. Even if the difference for subsequent ABG analysis was corrected by subtracting the gradient from the first ABG analysis, the ET-CO 2 to PaCO 2 difference varied from -1.3 to 1.7 kPa. The authors concluded that the efficacy of ET-CO 2 monitoring during OLV even when using corrected values remains questionable.

Two previous studies have evaluated noninvasive PCO 2 monitoring during OLV using both TC-CO 2 and ET-CO 2 devices in patients undergoing open thoracotomy,[5],[6] while there are no previous reports of using such monitoring in patients undergoing minimal access surgery. Oshibuchi et al compared the accuracy of TC-CO 2 and ET-CO 2 monitoring in a cohort of 26 adult patients undergoing OLV for open thoracotomy and pneumonectomy. The transcutaneous device (TCM3 transcutaneous CO 2 /oxygen device, Radiometer, Copenhagen, Denmark) was applied to the upper part of the patient's dependent arm. Evaluation of the TC-CO 2 to PaCO 2 difference revealed a bias of -0.4 mmHg and a precision of ± 2.5 mmHg during TLV and a bias of 1.4 mmHg and a precision of ± 4.3 mmHg during OLV. Evaluation of the ET-CO 2 to PaCO 2 difference revealed a bias of -5.8 mmHg and a precision of ± 4.1 mmHg during TLV and a bias of -7.1 mmHg and a precision of ± 4.6 mmHg during OLV. The authors concluded that TC-CO 2 monitoring was an accurate means of evaluating PaCO 2 during OLV. Tobias et al used the same transcutaneous device in their study of 15 young adult and pediatric patients (14.1 ± 6.1 years, range - 5 to 28 years) undergoing open thoracotomy.[5] During TLV, the TC-CO 2 to PaCO 2 difference was 2.5 ± 0.8 mmHg, while the ET-CO 2 to PaCO 2 difference was 3.9 ± 1.6 mmHg ( P = 0.0049). There was a significant increase in the ET-CO 2 to PaCO 2 gradient during OLV (5.8 ± 2.3 mmHg), while no change was noted in the TC-CO 2 to PaCO 2 difference (2.7 ± 1.4 mmHg).

The previously reviewed studies of Oshibuchi et al and Tobias, along with the data from the current cohort of adult patients, demonstrate the inaccuracy of ET-CO 2 monitoring during OLV and suggest that TC-CO 2 monitoring is an effective alternative or adjunct. Until recently, transcutaneous CO 2 monitoring was used most commonly in the neonatal and occasionally in the pediatric ICU population; however, there is growing experience with its use in adult patients in both the operating room and the ICU setting.[7],[8],[9],[14],[15],[16] These studies have demonstrated that TC-CO 2 is more accurate than ET-CO 2 in situations where the continuous monitoring of PaCO 2 is vital.

TC-CO 2 monitoring may be of particular benefit when the ventilation-perfusion properties of the respiratory system are altered. ET-CO 2 measures a sample of gas that contains a mixture of gas exhaled from several areas of the airway and alveoli. Regions with a high ventilation-perfusion ratio (dead space) do not participate in gas exchange and therefore the partial pressure of CO 2 is low or absent. During exhalation, the gas from regions of dead space mixes with the gas from areas of normal ventilation-perfusion ratios, resulting in dilution of the ET-CO 2 sample and a widening of the ET-CO 2 to PaCO 2 difference. Alternatively, areas of low ventilation-perfusion ratios (shunt) result in ineffective gas exchange and the addition of blood with a high partial pressure of CO 2 to the arterial circulation contributing to the increased ET-CO 2 to PaCO 2 gradient.[17],[18]

TC-CO 2 monitoring avoids the effect of sampling gas that may be subject to ventilation-perfusion mismatch. Transcutaneous monitoring relies on cutaneous respiration of the diffusion of gases across the skin. The transcutaneous monitor measures the CO 2 that is produced by local tissue metabolism and the CO 2 released from the blood as it flows through the capillaries near the skin surface. The latter is in direct equilibrium with the capillary CO 2 which is in equilibrium with the arterial CO 2 . Warming of the skin to 42°C by the sensor increases blood flow and CO 2 solubility, resulting in an even greater diffusion of CO 2 into the skin and equilibration with capillary and arterial PCO 2 values.[19],[20] The TC-CO 2 monitor measures the PCO 2 at the epidermis by using an infrared sensor, pH electrode or a Clark-type electrode.[21] Unlike ET-CO 2 , which typically underestimates actual CO 2 , the transcutaneous method typically overestimates actual CO 2 by 5.2-6.4 mmHg due to the increased CO 2 production from local metabolism induced by heating to 42°C. The currently available TC-CO 2 devices have an internal correction / calibration factor to correct for the heat-induced changes in CO 2 production. TC-CO 2 monitoring requires specific training in calibration, preparation, placement and maintenance of the device. Errors in any one of these steps may give false readings. When compared with ET-CO 2 monitoring, currently available TC-CO 2 monitors require a longer preparation time, including a 5-min calibration period and then an additional 5-10 min equilibration time after placement on the patient. Although not an issue with the TC-CO 2 monitor used in the current study, other TC-CO 2 monitors may require heating the skin to 44-45°C to ensure accuracy. When this is done, there are occasional reports of superficial burns and skin blistering. Technical and patient-related factors may affect the accuracy of TC-CO 2 monitoring. Improper membrane placement on the sensor or damage to the membrane may affect its accuracy. Patient factors, including skin thickness, skin edema and hypoperfusion (decreased cardiac output, hypovolemia or vasoconstriction), may also alter the diffusion of CO 2 to the sensor and result in inaccurate readings.[22],[23],[24] As no continuous noninvasive monitor can be expected to be 100% accurate, periodic calibration with an arterial sample may be indicated.

Conclusion

The current study adds to the growing body of knowledge demonstrating the efficacy of TC-CO 2 monitoring in the adult population. This study is the first to evaluate the use of TC-CO 2 vs. ET-CO 2 during OLV in minimal access surgery. The continuous monitoring of PaCO 2 may be particularly important during minimal access surgery. In addition to OLV, CO 2 insufflation, to facilitate surgical visualization, places patients at a higher risk of hypercarbia than those undergoing open procedures. The combination of increased CO 2 from systemic absorption of the insufflated CO 2 with alterations in tidal volume imposed by decreasing minute ventilation during OLV makes accurately and continuously monitoring PaCO 2 vital during minimal access surgery. TC-CO 2 monitoring is not meant to replace ET-CO 2 monitoring. Rather, the devices should be used to complement one another, especially in the OR setting. Although our data further demonstrate that TC-CO 2 monitoring is more accurate than ET-CO 2 monitoring, ET-CO 2 remains the standard of care in the OR to document the intratracheal position of the endotracheal tube, to serve as an additional ventilator disconnect monitor and to provide a capnograph for waveform analysis.

References

1.Bhavani-Shankar K, Moseley H, Kumar AY, Delph Y. Capnometry and anesthesia. Can J Anesth 1992;39:617-32.  Back to cited text no. 1  [PUBMED]  
2.Pansard JL, Chjolley B, Devilliers C, Clergue F, Viars P. Variation in arterial to end-tidal CO2 tension differences during anesthesia in the "kidney rest" lateral decubitus position. Anesth Analg 1992;75:506-10.  Back to cited text no. 2    
3.Grenier B, Verchere E, Meslie A, Dubreuil M, Siao D, Vandendriessche M, et al . Capnography monitoring during neurosurgery: Reliability in relation to various intraoperative positions. Anesth Analg 1999;88:43-8.  Back to cited text no. 3    
4.Ip Yam PC, Innes PA, Jackson M, Snowdon SL, Russell GN. Variation in the arterial to end-tidal PCO2 difference during one-lung thoracic anesthesia. Br J Anaesth 1994;72:21-4.  Back to cited text no. 4  [PUBMED]  
5.Tobias JD. Non-invasive carbon dioxide monitoring during one-lung ventilation: End-tidal versus transcutaneous techniques. J Cardiothor Vasc Anesth 2003;17:306-8.  Back to cited text no. 5  [PUBMED]  [FULLTEXT]
6.Oshibuchi O, Cho S, Hara T, Tomiyasu S, Makita T, Sumikawa K. A comparative evaluation of transcutaneous and dnd-tidal measurements of CO2 in thoracic anesthesia. Anesth Analg 2003;97:776-9.  Back to cited text no. 6    
7.Bhavani-Shankar K, Steinbrook RA, Mushlin PS, Freiberger D. Transcutaneous PCO2 monitoring during laparoscopic cholecystectomy in pregnancy. Can J Anaesth 1998;45:164-9.  Back to cited text no. 7  [PUBMED]  
8.Griffin J, Terry BE, Burton RK, Ray TL, Keller BP, Landrum AL, et al . Comparison of end-tidal and transcutaneous measures of carbon dioxide during general anesthesia in severely obese adults. Br J Anaesth 2003;91:498-501.  Back to cited text no. 8    
9.McBride DS Jr, Johnson JO, Tobias JD. Noninvasive carbon dioxide monitoring during neurosurgical procedures in adults: End-tidal versus transcutaneous techniques. South Med J 2002;95:870-4.  Back to cited text no. 9  [PUBMED]  
10.Eberhard P, Gisiger PA, Gardaz JP, Spahn DR. Combining transcutaneous blood gas measurement and pulse oximetry. Anesth Analg 2002;94:S76-80.  Back to cited text no. 10  [PUBMED]  
11.Rohling R, Biro P. Clinical investigation of a new combined pulse oximetry and carbon dioxide tension sensor in adult anesthesia. J Clin Monit 1999;15:23-7.  Back to cited text no. 11  [PUBMED]  [FULLTEXT]
12.Nunn JF, Hill DW. Respiratory dead space and arterial to end tidal CO2 tension difference in anesthetized man. J Appl Physiol 1960;15:383-9.  Back to cited text no. 12  [PUBMED]  [FULLTEXT]
13.Whitesell R, Asiddao C, Gollman D, Jablonski J. Relationship between arterial and peak expired carbon dioxide pressure during anesthesia and factors influencing the difference. Anesth Analg 1981;60:508-12.  Back to cited text no. 13  [PUBMED]  
14.Reid CW, Martineau RJ, Miller DR, Hull KA, Baines J, Sullivan PJ. A comparison of transcutaneous, end-tidal and arterial measurements of carbon dioxide during general anesthesia. Can J Anaesth 1992;39:31-6.  Back to cited text no. 14  [PUBMED]  
15.Tschupp A, Fanconi S. A combined ear sensor for pulse oximetry and carbon dioxide tension monitoring: Accuracy in critically ill children. Anesth Analg 2003;96:82-4.  Back to cited text no. 15  [PUBMED]  [FULLTEXT]
16.Bendjelid K, Schutz N, Stotz M, Gerard I, Suter PM, Romand JA. Transcutaneous PCO2 monitoring in critically ill adults: Clinical evaluation of a new sensor. Crit Care Med 2005;33:2203-6.  Back to cited text no. 16  [PUBMED]  [FULLTEXT]
17.Isert P. Control of carbon dioxide levels during neuroanesthesia: Current practice and an appraisal of our reliance upon capnography. Anaes Intens Care 1994;22:435-41.  Back to cited text no. 17  [PUBMED]  
18.Short JA, Paris ST, Booker PD, Fletcher R. Arterial to end-tidal carbon dioxide tension difference in children with congenital heart disease. Br J Anaesth 2001;86:349-53.  Back to cited text no. 18  [PUBMED]  [FULLTEXT]
19.Hansen TN, Sonoda Y, McIlroy MB. Transfer of oxygen, nitrogen and carbon dioxide through normal adult human skin. J Appl Physiol 1980;49:438-43.  Back to cited text no. 19  [PUBMED]  [FULLTEXT]
20.Hazinski TA, Severinghaus JW. Transcutaneous analysis of arterial PCO2. Med Instrum 1982;16:150-3.  Back to cited text no. 20  [PUBMED]  
21.Clark JS, Votteri B, Ariagno RL, Cheung P, Eichhorn JH, Fallat RJ, et al . Noninvasive assessment of blood gases. Am Rev Respir Dis 1992;145:220-32.  Back to cited text no. 21    
22.Eletr S, Jimison H, Ream AK, Dolan WM, Rosenthal MH. Cutaneous monitoring of systemic PCO2 on patients in the respiratory intensive care unit being weaned from the ventilator. Acta Anaesth Scand 1978;68:123-7.  Back to cited text no. 22  [PUBMED]  
23.Rithalia SV, Ng YN. Tinker J. Measurement of transcutaneous PCO2 in critically ill patients. Resuscitation 1982;10:13-8.  Back to cited text no. 23    
24.Rithalia SV, Clutton-Brock TH, Tinker J. Characteristics of transcutaeous carbon dioxide tension monitors in normal adults and critically ill patients. Intens Care Med 1984;10:149-53.  Back to cited text no. 24  [PUBMED]  

Copyright 2007 - Journal of Minimal Access Surgery


The following images related to this document are available:

Photo images

[ma07003f1.jpg] [ma07003t1.jpg] [ma07003f2.jpg]
Home Faq Resources Email Bioline
© Bioline International, 1989 - 2024, Site last up-dated on 01-Sep-2022.
Site created and maintained by the Reference Center on Environmental Information, CRIA, Brazil
System hosted by the Google Cloud Platform, GCP, Brazil