Indian Journal of Medical Microbiology
Medknow Publications on behalf of Indian Association of Medical Microbiology
ISSN: 0255-0857 EISSN: 1998-3646
Vol. 25, Num. 4, 2007, pp. 383-386

Indian Journal of Medical Microbiology, Vol. 25, No. 4, October-December, 2007, pp. 383-386

Original Article

A comparative study for the detection of mycobacteria by bactec MGIT 960, lowenstein jensen media and direct AFB smear examination

Department of Microbiology and Immunology, Sawai Man Singh Medical College, Jaipur - 302 004, Rajasthan

Correspondence Address: Department of Microbiology and Immunology, Sawai Man Singh Medical College, Jaipur - 302 004, Rajasthan, India. Email: drhemant@sancharnet.in.

Date of Submission: 06-May-2006
Date of Acceptance: 16-Jun-2007

Code Number: mb07104

Abstract

Purpose: To compare BACTEC MGIT 960 (M960) with conventional culture on Lowenstein Jensen (LJ) media and direct acid fast bacilli (AFB) smear examination for the detection of Mycobacteria in clinical samples obtained from suspected cases of pulmonary and extra pulmonary tuberculosis (TB).
Methods:
A total of 500 samples were processed for direct AFB smear examination, and culture on M960 and LJ media.
Results:
Two hundred fifty-eight out of 500 (51.6%) isolates of Mycobacteria were obtained by combined use of the two culture methods. Two hundred and fifty-three (50.6%) were positive in culture by M960 and LJ media and 28% (140/500) by direct AFB smear examination. The positivity rate of M960 system alone was 34.10% (88/258) and of LJ alone was 1.93% (5/258). Average time to detect growth (TTD) was 9.66 days by M960 and 28.81 days by LJ.
Conclusions:
M960 system is a rapid and sensitive method for early diagnosis of pulmonary and extrapulmonary TB. But for maximum recovery of Mycobacteria , a combination of both M960 and LJ media should be used.

Keywords: BACTEC MGIT 960, Lowenstein Jensen media

Tuberculosis (TB) is responsible for about one third of preventable deaths worldwide. [1] The spread of HIV/AIDS and emergence of multiple drug-resistant TB have further contributed to the worsening impact of the disease. [2] In India, 13 million people are infected and 3.5 million are positive for acid fast bacilli (AFB) with 2.2 million new TB cases being added every year. [3]

Although AFB microscopy and conventional Lowenstein Jensen (LJ) culture remain the cornerstone for the diagnosis of TB, the sensitivity of these traditional methods is quite low, especially in the samples containing small number of organisms. [4] There is a need for rapid, sensitive and accurate detection of these organisms in clinical specimens to hasten the administration of appropriate antimycobacterial therapy and prevent the spread of infection in the community. [5] A variety of manual and automated systems have been developed specifically to reduce the time to detect and identify Mycobacteria in clinical specimens. [5]

The present study was carried out to compare one such automated system, i.e., BACTEC MGIT 960 (M960) with conventional culture method, i.e., LJ media and direct AFB smear examination. M960 system is a fully automated, high capacity, non-radiometric, non-invasive instrument, which requires neither needles nor other sharp implements to incubate and monitor 960; 7 mL culture tubes. [6] The culture tubes contain Modified middle brook 7H9 media with fluorescent growth indicator embedded in silicone on the bottom of each tube. This compound is sensitive to the presence of dissolved oxygen in the broth. As the microorganisms grow in the media, oxygen gets depleted, allowing the fluorescence to be detected automatically over time. [5]

Materials and Methods

A total of 500 clinical samples were obtained from suspected cases of TB from different wards and out patient departments of the SMS and other allied hospitals of SMS Medical College, Jaipur, during the year 2004. The samples included sputum (330), cerebrospinal fluid (49), pleural fluid (41), bronchial washings (29), pus (17), urine (8), ascitic fluid (7), endometrium (5), ovarian cyst fluid (4), gastric aspirate (2), lymph node aspirate (2), bone scrapings (1), discharge from sinus (1), granulation tissue (1), menstrual blood (1), stool (1), and tissue from sinus tract (1).

A BBL MGIT tube (from Becton Dickinson) containing 7 mL modified middle brook 7H9 broth was used. Lyophilized MGIT PANTA (containing polymyxin B, azlocillin, nalidixic acid, trimethoprim, amphotericin B) was reconstituted with MGIT growth supplement (containing oleic acid, albumin, dextrose, catalase, polyoxyethylene stearate), and 0.8 mL of this was added prior to sample inoculation to the M960 tube. LJ media was prepared in the laboratory.

Processing of samples

Within 24 h of receiving the samples, direct smear was made, stained by Ziehl Neelsen method and graded as per RNTCP guidelines. For the purpose of digestion and decontamination, a maximum of 10 mL sample was taken in a falcon tube. An equal amount of a mixture containing 50 mL of 4% sodium hydroxide, 50 mL of 2.9% sodium citrate and 0.5% N -acetyl cysteine was added to it. The falcon tube was vortexed well and then kept at 37 °C for 20 min. After that, neutralization was done by filling the tube up to 50 mL mark with sterile phosphate buffered saline (pH 6.8). The tube was centrifuged for 20 min at 3000 x g . Supernatant was discarded and the pellet was reconstituted with 1 mL of sterile phosphate buffer, 0.5 mL of this was added to the M960 tube and two drops on the LJ media. Samples from sterile sites were inoculated as such or after centrifugation, without digestion and decontamination. The inoculated media were incubated at 37 °C, MGIT tubes were incubated inside the M960 instrument for 6 weeks and LJ media were incubated for 8 weeks and examined every week. M960 instrument detects the growth automatically; flashing red light to indicate instrument positive tubes and green for negative ones. Smears were made from M960-positive culture tubes as well as from M960 negative tubes that had some deposit in them, to confirm the presence or absence of Mycobacteria , as per the protocol of Becton Dickinson.

Results

Out of 500 samples processed, 50.6% (253/500) were positive by M960, 33.6% (165/500) by LJ media and 28% (140/500) by direct AFB smear examination. Total 51.6% (258/500) isolates of Mycobacteria were obtained by the combined use of these two culture methods. M960 was found to be positive in 98.06% (253/258) cases with average time to detect growth (TTD) around 9.66 days (2-39 days) and that of LJ 63.95% (165/258) with average TTD of 28.81 days (7-48 days). The positivity of direct AFB smear examination was found to be only 54.26% (140/258).

The positivity rate of M960 system alone was 34.10% (88/258) and of LJ alone 1.93% (5/258). Isolation rates of Mycobacteria in pulmonary and extrapulmonary samples were 61.83% (222/359) and 21.98% (31/141) by M960, 44.01% (158/359) and 4.96% (7/141) by LJ, respectively. One hundred and forty-two samples were from patients on anti-tubercular therapy. Out of these, 98.59% (140/142) were positive by M960, 69.01% (98/142) by LJ and 69.71% (99/142) by direct AFB smear examination. Details of isolation rates of Mycobacteria by M960 and LJ with respect to direct AFB smear results are shown in [Table - 1].

M960 instrument gave false-negative results for 5/258 (1.93%) samples. Out of these false negatives, two were tube positive but instrument negative. This was observed when smear was made from the deposit present in MGIT tube that had been declared negative by M960 instrument.

In our study, the break through contamination rate was found to be 13.4% (67/500) by M960 and 27.2% (136/500) by LJ. The break through contamination rate is defined as the detection of any instrument positive tube that is AFB smear negative and subculture negative yet Gram stain smear positive for other non acid fast microorganism. [6]

Discussion

The present study demonstrated that M960 system provided better isolation rate of Mycobacteria (98.06%) from a variety of clinical samples than the LJ media (63.95%). Various authors have reported similar findings ranging from 80 to 100% for M960 and from 59.7 to 87.2% for LJ. [5],[6],[7] In our study, isolation rate by M960 system was 34.11% more than that by LJ method. The additional positivity was seen both in pulmonary and extrapulmonary samples, not only in smear negative but in smear positive samples also. There was a low positivity rate shown by LJ method in our study in comparison to around 69-87.2% reported in literature. [6],[7] This could be because of the fact that samples that were grossly contaminated on LJ were considered negative, whereas in M960, since the smears were made from all instrument positive MGIT tubes, it was found that there were samples, which had both contaminants, as well as Mycobacteria grown in them. Such tubes were considered positive by M960.

Besides higher isolation rate, even the time to detect Mycobacteria was shorter on M960 than on LJ, average being 9.66 days (2-39) with M960 and 28.81 days (7-48) with LJ media. Similar findings have been reported in the literature. [8],[9],[10],[11],[12],[13] However in our study, time to detect Mycobacteria by M960 was lesser than that reported by others (11.6-14.4 days). [6],[7],[8],[10],[11],[13] This is probably because large number of samples (28%) were smear positive and incidence of TB is also higher in our country.

Time to detect Mycobacteria was directly related to the grade of positivity in direct AFB smear examination as seen in [Table - 2]. In smear positive samples, growth of Mycobacteria was detected in average 6.79 days (2-30) by M960.

The break through contamination rates in our study (13.4% for M960 and 27.2% for LJ) were higher in comparison to other studies (3.7-10% by M960 and 1.2-21.1% by LJ). [6],[8],[11],[12] This may be attributed firstly to the hot climatic conditions in our country and secondly to the longer transport time of the specimen to the laboratory in some cases, leading to overgrowth of the contaminants.

Thus, M960 was found to be most rapid and efficient system to isolate Mycobacteria . However, for maximum recovery of Mycobacteria , it is important to use both types of media as 1.93% isolates could be detected by LJ only.

As Mycobacteria can be isolated on an average within a week in smear positive samples, the antibiotic susceptibility tests can then be done and results would be available in another 4-13 days (as per BD protocol). This is in contrast to the time taken by LJ method, average 29 days for growth and another 30 days for sensitivity tests. This has great implications in monitoring the treatment failure and multidrug resistant TB patients. Though the cost of M960 culture system is approximately 8-10 times more than that for LJ media, the sensitivity and rapidity of the system are the major advantages that benefit not only the individual patient but also the community by controlling the disease and ultimately its transmission in the society.

References

1.Chitra C, Prasad CE. Evaluation of MGIT for primary isolation of Mycobacteria from clinical samples. Indian J Tuberc 2001; 48 :155-6.  Back to cited text no. 1    
2.Jesus de la Calle I, Jesus de la Calle MA, Rodriguez-Iglesis M. Evaluation of BDProbe Tec ET system as screening tool in direct detection of Mycobacterium tuberculosis complex in respiratory specimens. Diagn Microbiol Infect Dis 2003; 47 :573-8.  Back to cited text no. 2    
3.Oberoi A, Kaur H. Comparison of rapid colorimetric method with conventional method in the isolation of Mycobacterium tuberculosis . Indian J Med Microbiol 2004; 22 :44-6.  Back to cited text no. 3  [PUBMED]  [FULLTEXT]
4.Negi SS, Khan SF, Gupta S, Pasha ST, Khare S, Lal S. Comparison of conventional diagnostic modalities, BACTEC culture and PCR test for diagnosis of tuberculosis. Indian J Med Microbiol 2005; 23 :29-33.  Back to cited text no. 4  [PUBMED]  [FULLTEXT]
5.Lu D, Heeren B, Dunne WM. Comparison of the automated MGIT with Lowenstein Jensen media for recovery of Mycobacteria from clinical specimens. Am J Clin Pathol 2002; 118 :542-5.  Back to cited text no. 5  [PUBMED]  [FULLTEXT]
6.Hanna BA, Ebrahimzadeh A, Elliot LB, Morgan MA, Novak SM, Rusch-Gerdes S, et al . Multicentre evaluation of BACTEC MGIT 960 system for recovery of Mycobacteria. J Clin Microbiol 1999; 37 :748-52.  Back to cited text no. 6    
7.Alcaide F, Benitez MA, Escriba JM, Martin R. Evaluation of BACTEC MGIT 960 and the MB/BacT systems for recovery of Mycobacteria from clinical specimens and for species identification by DNA AccuProbe. J Clin Microbiol 2000; 38 :398-401.  Back to cited text no. 7    
8.Tortoli E, Cichero P, Piersimoni C, Simonetti MT, Gesu G, Nista D. Use of BACTEC MGIT 960 for recovery of Mycobacteria from clinical specimens: Multicentre study. J Clin Microbiol 1999; 37 :3578-82.  Back to cited text no. 8  [PUBMED]  [FULLTEXT]
9.Kanchana MV, Cheke D, Natyashak I, Connor B, Warner A, Martin T. Evaluation of BACTEC MGIT 960 system for the recovery of Mycobacteria . Diagn Microbiol Infect Dis 2000; 37 :31-6.  Back to cited text no. 9    
10.Idiogras P, Beristain X, Iturzaeta A, Vicente D, Perez Trallero E. Comparison of the automated non radiometric BACTEC MGIT 960 system with Lowenstein Jensen media, Colestos, MB 7H11 solid media for recovery of Mycobacteria. Eur J Clin Microbiol Infect Dis 2000; 19 :350-4.  Back to cited text no. 10    
11.Somoskovi A, Kodman C, Lantos A, Bartfai Z, Tamasi L, Fuzy J, et al . Comparison of recoveries of Mycobacterium tuberculosis using the automated BACTEC MGIT 960system, BACTEC 460 TB system and Lowenstein Jensen media . J Clin Microbiol 2000; 38 :2395-7.  Back to cited text no. 11    
12.Chien HP, Yu MC, Wu MH, Lin TP, Luh KT. Comparison of the BACTEC MGIT 960 with Lowenstein Jensen media for recovery of Mycobacteria from clinical specimens. Int J Tuberc Lung Dis 2000; 4 :866-70.  Back to cited text no. 12  [PUBMED]  [FULLTEXT]
13.Huang TS, Chen CS, Lee SS, Huang WK, Liu YC. Comparison of BACTEC MGIT 960 and BACTEC 460 TB system for the detection of Mycobacteria in clinical specimens. Ann Clin Lab Sci 2001; 31 :279-83.  Back to cited text no. 13  [PUBMED]  [FULLTEXT]

Copyright 2007 - Indian Journal of Medical Microbiology


The following images related to this document are available:

Photo images

[mb07104t2.jpg] [mb07104t1.jpg]
Home Faq Resources Email Bioline
© Bioline International, 1989 - 2024, Site last up-dated on 01-Sep-2022.
Site created and maintained by the Reference Center on Environmental Information, CRIA, Brazil
System hosted by the Google Cloud Platform, GCP, Brazil